Re: Salt testing

From: Robert A. LaBudde (ral@lcfltd.com)
Date: Fri Jun 29 2001 - 12:37:12 PDT

  • Next message: Ken Hilderbrand: "RE: Frozen tuna loins"

    At 11:50 AM 6/29/01 -0400, Mike wrote:
    >We are currently using Quantab titrator strips for chloride for our salt
    >analysis to determine %water phase salt in product.
    >
    >While the strips work OK, I am looking for a better accuracy in
    >determining salt content and % water phase salt.
    >
    >However, we would like to keep costs and time expenditures low, and also
    >stay away from wet chemistry procedures (we don't really have a
    >chemistry lab). Kind of looks like we want both ends and the middle
    >here, but I may as well go for it.

    The least expensive and reliable method for salt in this type of material
    is via conductivity measurement. Brine content would obviously require a
    moisture analysis (typically microwave for rapid, drying oven for slow).

    The Quantab system is specific for chloride (or other silver precipitating
    anions), while conductivity is non-specific, and would be influenced by
    phosphates, for example. Conductivity measures "total salts", so you will
    have a slight high bias vs standard chloride analysis.

    The Quantab system is very accurate on pure salt solutions when the test
    kit is fresh. It deteriorates over time, either in the bottle or from
    reopening the bottle over shorter times. So unless you have a high rate of
    turnover of bottles (e.g., every 2 weeks), the Quantab system may be
    unreliable. It is for this reason that I no longer recommend it for salt
    analysis of foods.

    Sample preparation is of some importance: You need to extract the salt from
    the muscle adequately. Cold water with a 10x - 25x dilution will generally
    work in a blender. If you're using Quantabs, make sure you filter the
    extract through a coarse filter (fast qualitative paper or a coffee
    filter), otherwise you may impede uptake into the test strip.

    Other rapid methods include ion-selective electrodes, such as chloride or
    sodium to replace the measurement method.

    Typically conductivity measurements result in an precision of about 0.2%
    standard deviation and a bias < 0.3% relative to the Volhard standard method.

    You should keep the dilution rate high (e.g., 25x) and the conductivity low
    (e.g., < 2000 uS) in order to keep the conductivity response linear.

    Let me know if you are interested in obtaining a method and equipment for
    conductivity salt testing, email me privately and I can supply sources.

    ================================================================
    Robert A. LaBudde, PhD, PAS, Dpl. ACAFS e-mail: ral@lcfltd.com
    Least Cost Formulations, Ltd. URL: http://lcfltd.com/
    824 Timberlake Drive Tel: 757-467-0954
    Virginia Beach, VA 23464-3239 Fax: 757-467-2947

    "Vere scire est per causas scire"
    ================================================================



    This archive was generated by hypermail 2b29 : Fri Jun 29 2001 - 12:38:01 PDT